pH : |
|
|
Readout/ Resolution : |
-2.000 to 19.999 or 0 to 14 in aqueous solutions /
0.1,0.01,0.001 |
|
Relative Accuracy : |
±
0.002 |
|
SMART STABILITY™ and SMART AVERAGING™ : |
Resolution dependent / automatic |
|
mV : |
|
|
Range – mV / RmV : |
±
1999.9 |
|
ORP : |
EH
autocalibration to standard hydrogen electrode |
|
Resolution : |
0.1 |
|
Relative Accuracy : |
±
0.2 mV or 0.05%, whichever is greater |
|
CONDUCTIVITY : |
|
|
Range : |
0.000 to 3000 mS/cm, auto resolution with cell constant
dependence |
|
Resolution : |
4
significant digits down to 0.001 µS/cm, cell constant
dependent |
|
Relative Accuracy : |
0.5% ± 1 digit or 0.01 µS/cm, whichever is greater |
|
Cell Constant : |
0.001 to 199.9 |
|
Cell Compatibility : |
2
Cell, 4 Cell, Planar Cell |
|
SMART STABILITY™ and SMART AVERAGING™ : |
Resolution dependent / automatic |
|
DirectCal™ / AUTO-CAL™ : |
1
to 5 points / 1 to 3 points |
|
RESISTIVITY : |
|
|
Range / Resolution : |
0.0001 to 100 Megohm / Auto resolution |
|
Relative Accuracy : |
0.5 % ± 1 digit |
|
SALINITY : |
|
|
Range / Resolution : |
0.01 to 80.0 ppt NaCl equivalent, 0.01 to 42ppt
practical salinity / 0.01 |
|
Relative Accuracy : |
±
0.1 ± 1 digit |
|
TDS : |
|
|
Range / Resolution : |
0
to 19999 / 1 mg/L |
|
Relative Accuracy : |
±
0.5 % ± 1 digit |
|
Adjustable TDS Factor : |
0.00 to 10.00 |
|
TEMPERATURE : |
|
|
Range / Resolution : |
-5
to 105 °C / 0.1 up to 99.9 °C, 1.0 over 99.9 °C |
|
Relative Accuracy : |
±
0.1 °C |
|
Temperature Compensation : |
Cond (nLF, linear 0.0 to 10.0% / °C); pH (auto/manual) |
|
Display : |
Custom backlit LCD, for easy viewing |
|
INPUTS / OUTPUTS : |
|
|
Sensor Inputs : |
Conductivity-miniDIN - Locking; ATC-miniDIN-Locking; pH
- BNC; Reference-Pin tip; Stirrer-Phono-jack |
|
Solid State pH Sensor : |
ISFET |
|
Communications : |
Bi-directional RS232 - Stereo jack, optional USB
interface Cable |
|
SPECIAL GLP SOFTWARE FEATURES : |
|
|
# Of Data Logging Points : |
200 points by time, stability or manual read with
time/date stamp |
|
Calibration Log : |
Last 10 Calibrations |
|
Advanced Self-test and Diagnostics : |
YES |
|
GLP Electrode Verification : |
YES |
|
Password Protection : |
YES |
|
Internet Software Updates / Downloadable Methods : |
YES |
|
21 CFR Part 11 Software Package : |
Optional |
|
Power : |
Universal power adapter and battery power 4 x AA
batteries |
|
Regulatory and Safety : |
CE, CSA, UL, TÜV, FCC Class A limits |
|
ENVIRONMENTAL OPERATING CONDITIONS : |
|
|
Ambient Operating Temperature : |
5
to 45 °C |
|
Relative Humidity : |
5
to 85% Non-Condensing |
|
IP Rating : |
IP54, splashproof and dustproof |
|
Warranty : |
36
months (from date of purchase) |
|
* Above data generated under optimal/ controlled
circumstances |