This waterproof, battery-operated pH/ISE meter stands up
to the most demanding quality control and general analysis
applications and delivers over 2,000 hours of continuous
operation. Convenient Quick-Start Guide is located on the
back of the meter to make the unit easy to operate in the
field.
The rugged meter provides reliable measurement of pH and
ionic concentrations and offers a full complement of GLP-related
features—including available 21 CFR Part 11 software option.
Automatic calibration or 1 to 5 point manual calibration for
pH and direct ion measurement ensures accuracy even at
extremes of concentration.
SMART STABILITYTM and SMART AVERAGINGTM functions simplify
the measurement process. You can record and store up to 10
setup parameters and calibrations.
Additional Features:
IP67 waterproof (may be immersed for up to one hour with no
water incursion)
Waterproof BNC for either pH or ISE electrode
Automatic or manual pH calibration with 1 to 5 calibration
points
Recognizes NIST, US and manual buffers
Manual ISE calibration with 1 to 5 calibration points
Point-to-point ISE calibration, high and low concentration
range
Automatic blank correction
Self-diagnostics
GLP IQ/OQ included
200 point datalogger which is downloadable to printer or
computer via RS232 or USB accessories
CE, CSA, UL, TÜV and FCC Class A limits approved
Three-year replacement warranty.
|
pH : |
|
|
Readout Range / Resolution : |
-2.000 to 19.999 or 0 to 14 in aqueous solutions /
0.1,0.01,0.001 |
|
Relative Accuracy : |
±
0.002 |
|
SMART STABILITY™ and SMART AVERAGING™ : |
Resolution dependent / automatic |
|
AUTO-CAL™ and/or Manual Calibration : |
1
to 5 points; US/NIST, DIN, and custom buffers recognized
|
|
lSE : |
|
|
Range / Resolution : |
0
to 19900 / 1,2, or 3 significant digits |
|
Relative Accuracy : |
±
0.2 mV or 0.05%, whichever is greater |
|
SMART STABILITY™ and SMART AVERAGING™ : |
Resolution dependent / automatic |
|
Units : |
M,
mg/L, %, ppb, or no units |
|
# ISE Cal points : |
1
to 5 |
|
Special ISE Calibration Features : |
Linear point to point, selectable auto-blank non-linear,
low concentration range |
|
mV : |
|
|
Range – mV / RmV : |
±
1999.9 |
|
ORP : |
EH
autocalibration to standard hydrogen electrode |
|
Resolution : |
0.1 |
|
Relative Accuracy : |
±
0.2 mV or 0.05%, whichever is greater |
|
TEMPERATURE : |
|
|
Range / Resolution : |
-5
to 105 °C / 0.1 up to 99.9 °C, 1.0 over 99.9 °C |
|
Relative Accuracy : |
±
0.1 °C |
|
Temperature Compensation : |
Auto/Manual |
|
Display : |
Custom backlit LCD, for easy viewing |
|
INPUTS / OUTPUTS : |
|
|
Sensor Inputs : |
pH/ISE - BNC; ATC-miniDIN-Locking; Reference-Pin tip
(all waterproof) |
|
Communications : |
Waterproof bi-directional RS232 - Stereo jack, optional
USB interface Cable |
|
SPECIAL GLP SOFTWARE FEATURES : |
|
|
# Of Data Logging Points : |
200 points by time, stability or manual read with
time/date stamp |
|
Calibration Log : |
Last 10 Calibrations |
|
User Method Storage : |
10
Saved |
|
Advanced Self-test and Diagnostics : |
YES |
|
GLP Electrode Verification : |
YES |
|
Password Protection : |
YES |
|
Internet Software Updates / Downloadable Methods : |
YES |
|
21 CFR Part 11 Software Package : |
Optional |
|
Power : |
4
x AA batteries, powerless date protection |
|
Regulatory and Safety : |
CE, CSA, UL, TÜV, FCC Class A limits |
|
ENVIRONMENTAL OPERATING CONDITIONS : |
|
|
Ambient Operating Temperature : |
5
to 45 °C |
|
Relative Humidity : |
5
to 85% Non-Condensing |
|
IP Rating : |
IP67, waterproof and dustproof |
|
Warranty : |
36
months (from date of purchase) |
|
*Above data generated under optimal/controlled